“SLIME” on the probe cover provides better-quality ultrasound images without water-soluble gel

To the Editor,

While performing ultrasonography for ultrasound-guided nerve block, an ultrasound transmission gel (water-soluble gel) is used to obtain good-quality ultrasound images. Air entering between the probe and probe cover interferes with the quality of the ultrasound images; however, it is sometimes difficult for beginners to prevent entry of air underneath the probe cover. On the other hand, the use of gel makes the probe sticky after the procedure. While a directly glued ultrasound probe cover without the use of gel has been reported [1], it is expensive.

We have found that the use of so-called “SLIME” (a mixture of polyvinyl alcohol and borax) is useful to obtain ultrasound images equivalent in quality to those obtained using gel.

We applied SLIME directly on a linear probe (S-Nerve HFL50x/15-6®, FUJIFILM Medical Co., Ltd., Tokyo, Japan) (Fig. 1a) and covered the probe and SLIME with a probe cover. Simple application of SLIME allows it to be placed firmly on the probe, and it is easy even for beginners to eliminate entry of air when it is covered with the probe cover (Fig. 1b). The quality of the ultrasound image using SLIME is equivalent to using gel (Fig. 1c, d). Preparation of SLIME is easy [2], as it is just a simple mixture of polyvinyl alcohol (included in laundry glue, KANEYO NOL®, KANEYO SOAP Co., Ltd., Tokyo, Japan) and borax (Housya®, KENEI Pharmaceutical Co., Ltd., Osaka, Japan), both of which are commercially available. The method of preparation of SLIME is as follows. First, mix equal amounts of laundry glue and water (50 ml each). Second, prepare the aqueous solution of borax with 2 g of borax in 25 ml of water. Third, pour this supernatant borax aqueous solution little by little into the water suspension of laundry glue and mix until the mixture becomes hard [2]. Although SLIME contains air, micro air might not interfere with the quality of ultrasound images. Subsequent to the ultrasound procedure, the SLIME can be peeled off easily from the probe without sticking which is different to gel (Fig. 1e, f) and is reusable. It is safe, unless the borax comes in direct contact with the skin.

Fig. 1
figure 1

a Simple application of SLIME on the probe allows the SLIME to be placed firmly on the probe. b Application of SLIME on the probe allows entry of air between the probe and probe cover to be avoided. c The ultrasound image of the median nerve at the forearm obtained with the use of SLIME on the probe. d The ultrasound image with the use of water-soluble gel. e The probe remains smooth after removal of the probe cover using SLIME. f The probe after removal of the probe cover using gel

Conclusions

The ultrasound image quality using SLIME is as equivalent as to that using water-soluble gel. In addition, SLIME as the probe cover might be easy even for beginners to eliminate entry of air and to treat the probe after the procedure.

Availability of data and materials

Not applicable.

Abbreviations

SLIME:

A mixture of polyvinyl alcohol and borax

References

  1. Marhofer P, Fritsch G. Sterile working in ultrasonography: the use of dedicated ultrasound covers and sterile ultrasound gel. Expert Rev Med Devices. 2015;12:667–73.

    Article  Google Scholar 

  2. Let’s make a slime (AS ONE Corporation 2017). https://axel.as-1.co.jp/contents/t_tb/create_slime. Accessed 1 Apr 2019.

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

YO prepared the SLIME, performed the experiments, wrote the manuscript, responded to the critical comments, and obtained the final approval.

Corresponding author

Correspondence to Yoshimune Osaka.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

Not applicable.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Osaka, Y. “SLIME” on the probe cover provides better-quality ultrasound images without water-soluble gel. JA Clin Rep 5, 43 (2019). https://doi.org/10.1186/s40981-019-0263-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40981-019-0263-9