Right ventricular functional assessment by three-dimensional transesophageal echocardiography is useful for withdrawal from a right ventricular assist device: a case report

Abstract

Right ventricular assist device (RVAD) implantation is one type of surgical treatment used for right heart failure. It is important to assess right ventricular (RV) function precisely when RVAD withdrawal is considered. Although assessment of RV function is difficult due to its complicated shape and contraction pattern, the volumetric analysis method of three-dimensional (3D) transesophageal echocardiography (TEE) has been developed and is useful for this task. We report the case of a 79-year-old man who successfully underwent RVAD withdrawal and evaluation using 3D TEE. 3D TEE had an important role in determining the timing of withdrawal from RVAD in this case.

Background

Temporary right ventricular assist device (RVAD) implantation is an acceptable treatment for managing postoperative right ventricular (RV) failure in severely ill patients after cardiac surgery and left ventricular assist device (LVAD) recipients [1, 2]. Because criteria for RVAD withdrawal have not yet been established, it is important to assess RV function to determine whether RVAD can be removed. Although conventional two-dimensional (2D) echocardiography (2DE) is often performed to assess RV function, there are limits when measuring complicated RV [3, 4]. We report a case of RV functional assessment using three-dimensional (3D) echocardiography (3DE) for withdrawal from RVAD.

Case presentation

A 79-year-old man (158 cm/59 kg) was diagnosed with having aortic valve stenosis and followed-up. Although he was asymptomatic, transthoracic echocardiography showed progression of stenosis (aortic valve area of 0.76 cm2, mean pressure gradient of 58 mmHg, peak pressure gradient of 89 mmHg, left ventricular ejection fraction of 69%). Aortic valve replacement and pulmonary vein isolation for paroxysmal atrial fibrillation were scheduled. During the procedure, the right coronary artery base was damaged and acute right heart failure developed. Although right coronary bypass grafting (aorta-saphenous vein–right coronary artery) was performed, RV function did not improve. Therefore, RVAD was implanted. Assessment of RV function was attempted using transesophageal echocardiography (TEE) on postoperative day (POD) 15, as requested by the cardiovascular surgeon. During examination, TEE revealed tricuspid annular plane systolic excursion (TAPSE) of 7 mm and RV fractional area change (RVFAC) of 31%. We also measured right ventricular ejection fraction (RVEF) using newly developed software (4D RV-FUNCTION®; TomTec Imaging Systems, Unterschleißheim, Germany), which yielded a result of 23.7% (Fig. 1). TAPSE, RVFAC, and RVEF according to 3D analysis were much lower than the reference volume. Therefore, we concluded that RVAD withdrawal would be difficult. Because the request was made again on POD 28, the right heart function was evaluated once more by TEE. Although RV function was still poor, as demonstrated with TAPSE of 6.6 mm and RVFAC of 32%, RVEF as measured using 3D analysis had improved to 34.9% (Fig. 2), which was close to the low reference value of RVEF. The 2D and 3D measurements were performed with the RVAD flow rate decreased to less than 0.5 L/min, continuous administration of dobutamine 5 μg/kg/min, and decreased mechanical ventilation. Pulmonary hypertension was not observed during measurements. Measurement by TEE was performed only two times because the recovery of RV function took some time. Based on TEE findings, we considered that it was possible to withdraw RVAD with increased inotropes. RVAD was successfully removed on POD 29. Hemodynamics was stable after withdrawal of RVAD. The patient was discharged from the intensive care unit on POD 39 and from the hospital on POD 110. The patient has been well since then.

Fig. 1
figure 1

Three-dimensional right ventricular ejection fraction on postoperative day 15. Measurements were performed using 4D RV-FUNCTION® (TomTec Imaging Systems), which yielded a result of 23.7%

Fig. 2
figure 2

Three-dimensional right ventricular ejection fraction on postoperative day 28. Right ventricular ejection fraction evaluated by three-dimensional analysis improved to 34.9%

Discussion

Implantation of a temporary RVAD has recently been shown to be of value for managing postoperative RV failure. However, reports to date have been limited to small series of patients or isolated case reports, and withdrawal criteria have not yet been established [1, 2]. The course of VAD is complicated by high rates of major adverse events such as thromboembolism, device infections, and mechanical complications [2]. In this context, because it is essential to determine when RVAD can be removed due to the recovery of the right side of the patient’s heart, precise RV function assessment is important. Hemodynamics and echocardiography may provide useful information, but often they are not sufficient [5]. Although it has been reported that the filling characteristics of LVAD provide information that is useful to determine whether a patient can be successfully weaned from RVAD, these may not be applicable for all patients because the LVAD type has been recently changed from pulsatile to continuous flow [5].

Many investigations suggest that RV function plays an important role in the morbidity and mortality of patients presenting with signs and symptoms of cardiopulmonary diseases [6,7,8,9]. Therefore, it is important to assess RV function. Evaluations of RV volume, function, and mass are challenging because of the geometrical complexity and individual differences.

There are various measurement methods for RV function, such as magnetic resonance imaging (MRI), echocardiography, and the thermodilution method, which uses a pulmonary artery catheter. Resolution and analysis methods are advancing and have become useful for the assessment of cardiac function [3, 7, 10, 11]. RV systolic function has been evaluated with several parameters, namely, RV index of myocardial performance, TAPSE, RVFAC, tissue Doppler–derived tricuspid lateral annular systolic velocity (S′), and longitudinal strain and strain rate (J). RV contraction comprises the following: (1) movement of the RV free wall toward the interventricular septum; (2) projection of the interventricular septum toward RV with curvature radius reduction of the interventricular septum during left ventricular contraction; (3) traction of the RV free wall with left ventricular contraction at the holdfast of the interventricular septum; and (4) RV contraction in the long axis direction [12]. RV has a complicated shape and contracts in three dimensions. Furthermore, the movement is affected by the interventricular septum and left ventricle. Therefore, it is difficult to assess the volume and ejection fraction of RV in two dimensions. In addition, it has been reported that the accuracy of 2D analysis methods such as TAPSE and S′ are insufficient compared with that of MRI [4]. 3DE allows direct measurement of the RV volume without relying on geometric assumptions regarding RV shape, resulting in more accurate and reproducible measurements of RVEF. In addition, 2D measurements are inaccurate in cases of arrhythmia or impaired diastolic dysfunction, whereas 3D methods allow measurement of RV function in these cases [3, 7, 10, 11, 13]. Nonetheless, RVFAC is correlated with RVEF. Although 2D measurements were not able to capture the recovery of RV function, 3D measurements did in our case. RVEF of 44% or more is considered normal [7], but it was decreased in this case. Quantitative analysis was performed using newly dedicated software for our case (4D RV Function 2.0; TomTec Imaging Systems), which provided measurement values of RV volumes and EF using the speckle-tracking technology. Compared with MRI, the new software is fast, reproducible, and accurate over a wide range of RV sizes and functions [11]. It allowed us to perform rapid assessments and led to RVAD withdrawal. Although several reports showed that TEE predicted successful withdrawal of VAD [5, 14], they have not mentioned about 3D TEE. This may provide more accurate evaluations regarding withdrawal than does 2D TEE.

Conclusions

Three-dimensional TEE had an important role in determining the timing of withdrawal from RVAD in this case.

Abbreviations

2D:

Two-dimensional

2DE:

Two-dimensional echocardiography

3D:

Three-dimensional

3DE:

Three-dimensional echocardiography

EF:

Ejection fraction

LVAD:

Left ventricular assist device

MRI:

Magnetic resonance imaging

POD:

Postoperative day

RV:

Right ventricular

RVAD:

Right ventricular assist device

RVEF:

Right ventricular ejection fraction

RVFAC:

RV fractional area change

TAPSE:

Tricuspid annular plane systolic excursion

TEE:

Transesophageal echocardiography

VAD:

Ventricular assist device

References

  1. Kaul T. Postoperative acute refractory right ventricular failure: incidence, pathogenesis, management and prognosis. Cardiovasc Surg. 2000;8:1–9.

    Article  CAS  PubMed  Google Scholar 

  2. Aissaoui N, Morshuis M, Schoenbrodt M, Meibodi KH, Kizner L, Börgermann J, Gummert J. Temporary right ventricular mechanical circulatory support for the management of right ventricular failure in critically ill patients. J Thorac Cardiovasc Surg. 2013;146:186–91.

    Article  PubMed  Google Scholar 

  3. Zwaan HB, Geleijnse ML, Mcghie JS, Boersma E, Helbing WA, Meijboom FJ, Roos-Hesselink JW. Right ventricular quantification in clinical practice: two-dimensional vs. three-dimensional echocardiography compared with cardiac magnetic resonance imaging. Eur J Echocardiogr. 2011;12:656–4.

    Article  PubMed  Google Scholar 

  4. Ling LF, Obuchowski NA, Rodriguez L, Popovic Z, Kwon D, Marwick TH. Accuracy and interobserver concordance of echocardiographic assessment of right ventricular size and systolic function: a quality control exercise. J Am Soc Echocardiogr. 2012;25:709–13.

    Article  PubMed  Google Scholar 

  5. Mandarino WA, Winowich S, Gasior TA, Pham S, Griffith BP, Konnos RL. Assessment of timing right ventricular assist device withdrawal using left ventricular assist device filling characteristics. ASAIO J. 1997;43:801–5.

    Article  Google Scholar 

  6. Bleasdale RA. Prognostic importance of right ventricular dysfunction. Heart. 2002;88:323–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. Rudski LG, Lai WW, Afilalo J, Hua L, Handschumacher MD, Chandrasekaran K, Solomon SD, Louie EK, Schiller NB. Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography. J Am Soc Echocardiogr. 2010;23:685–713.

    Article  PubMed  Google Scholar 

  8. Mehta SR, Eikelboom JW, Natarajan MK, Diaz R, Yi C, Gibbons RJ, Yusuf S. Impact of right ventricular involvement on mortality and morbidity in patients with inferior myocardial infarction. J Am Coll Cardiol. 2001;37:37–43.

    Article  CAS  PubMed  Google Scholar 

  9. Sabe MA, Sabe SA, Kusunose K, Flamm SD, Griffin BP, Kwon DH. Predictors and prognostic significance of right ventricular ejection fraction in patients with ischemic cardiomyopathy clinical perspective. Circulation. 2016;134:656–65.

    Article  PubMed  Google Scholar 

  10. Shiota T. 3D echocardiography: evaluation of the right ventricle. Curr Opin Cardiol. 2009;24:410–4.

    Article  PubMed  Google Scholar 

  11. Medvedofsky D, Addetia K, Patel AR, Sedlmeier A, Baumann R, Mor-Avi V, Lang RM. Novel approach to three-dimensional echocardiographic quantification of right ventricular volumes and function from focused views. J Am Soc Echocardiogr. 2015;28:1222–31.

    Article  PubMed  Google Scholar 

  12. Sheehan F, Redington A. The right ventricle: anatomy, physiology and clinical imaging. Heart. 2008;94:1510–5.

    Article  PubMed  Google Scholar 

  13. Imada T, Kamibayashi T, Ota C, Shibata SC, Iritakenishi T, Sawa Y, Fujino Y. Intraoperative right ventricular fractional area change is a good indicator of right ventricular contractility: a retrospective comparison using two- and three-dimensional echocardiography. J Cardiothorac Vasc Anesth. 2015;29:831–5.

    Article  PubMed  Google Scholar 

  14. Barzilai B, Dávila-Román VG, Eaton MH, Rosenbloom M, Spray TL, Wareing TH, Cox JL, Kouchoukos NT. Transesophageal echocardiography predicts successful withdrawal of ventricular assist devices. J Thorac Cardiovasc Surg. 1992;104:1410–6.

    CAS  PubMed  Google Scholar 

Download references

Acknowledgements

We thank our patient for providing permission to publish this case report.

Funding

The authors declare that they have no funding.

Author information

Authors and Affiliations

Authors

Contributions

TH wrote the main paper with advice from IT and FY. All authors discussed the case presentation and discussion and commented on the manuscript at all stages. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Hiroki Taenaka.

Ethics declarations

Ethics approval and consent to participate

Written informed consent was obtained from all subjects for publication of this case report and accompanying images.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Taenaka, H., Imada, T., Abe, R. et al. Right ventricular functional assessment by three-dimensional transesophageal echocardiography is useful for withdrawal from a right ventricular assist device: a case report. JA Clin Rep 3, 40 (2017). https://doi.org/10.1186/s40981-017-0112-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40981-017-0112-7

Keywords