From: Using the Bair Hugger™ temperature monitoring system in neck and chest regions: a pilot study

Fig. 1

Pearson’s correlation between esophageal temperature and other temperatures. a Plots of temperature data obtained from the forehead (THead) and esophagus (TEso), b neck (TNeck) and TEso, and c anterior chest (TChest) and TEso. Thin lines in the figure denote the line of identify. Pearson’s correlation coefficient indicated a strong correlation for all comparisons (r = 0.81 between TEso and THead, r = 0.86 between TEso and TNeck, and r = 0.84 between TEso and TChest)