Difficult ventilation due to an undetected mask crack

Letter to Editor

We present a case that experienced difficulties in ventilation due to an undetected crack in the anesthesia mask. This report underscores the significance of comprehensive pre-anesthesia equipment inspections, including a meticulous assessment of the mask, to uphold patient safety.

A 72-year-old male (height 177 cm, weight 83 kg), ASA-PS II, was scheduled for thoracoscopic lobectomy. Preoperatively, the anesthesia machine (Aespire, Datex-Ohmeda) and circuit were checked for leaks [1]. The patient was preoxygenated with 6 L 100% O2 using a mask (Laerdal silicone mask, without a multifunction mask cover). Anesthesia was induced with 100 mg propofol, 0.3 μg/kg/min remifentanil, and 70 mg rocuronium. Capnography waveforms demonstrated the absence of a plateau phase, indicating abnormal mask ventilation (V2 level) after induction [2]. Despite the insertion of an oral airway, ventilation became increasingly challenging, leading to the complete disappearance of capnography waveforms (V3 level) and a call for emergency backup [2].

The responding anesthesiologist, feeling discomfort in his left hand holding the mask, identified a crack in the mask (Fig. 1), promptly replaced it, and restored effective mask ventilation, presenting the V1 waveform on capnography. During the period from preoxygenation until intubation, the patient's minimum percutaneous oxygen saturation reached 78%.

Fig. 1
figure 1

Anesthesia mask with crack

According to the 2014 JSA Airway Management Guidelines [2], immediate intubation is recommended for patients in V2 or V3 status. In hindsight, we believe intubation attempts should have been initiated earlier, concurrent with the call for additional support.

First reported mask breakage in our OR. We posit that the ventilation challenge stemmed from the unnoticed mask crack, as ventilation only improved after mask replacement. It is likely that the crack was present prior to the preoxygenation phase, given that the capnography waveform did not exhibit the V1 reading until the mask was replaced. The minute size of the crack may have limited airway pressurization, resulting in partial ventilation. Unfortunately, efforts to manually adjust mask pressure inadvertently exacerbated the crack, rendering airway pressure maintenance impossible. The exact duration of use for this mask remains uncertain due to incomplete records. Upon reviewing a photo of the damaged mask, the manufacturer noted discoloration, indicating potential silicone deterioration that might have contributed to the cracks, highlighting an increased risk of breakage.

In this case, the pre-anesthesia equipment examination adhered to the “Checkout Procedures of Anesthesia Apparatus” [1], but did not encompass a detailed mask inspection. We recommend an extension of the current protocol to include such an assessment, which could have potentially prevented this incident.

This case underscores the critical importance of meticulously inspecting the anesthesia mask during pre-operative checks. Anesthesiologists must receive education regarding their pivotal role in safeguarding patient well-being.

Availability of data and materials

Not applicable.

References

  1. Japanese Society of Anesthesiologists: Checkout Procedures of Anesthesia Apparatus ver.6–3 (Japanese). https://anesth.or.jp/files/pdf/guideline_checkout_20221117.pdf. Accessed 26 May 2023.

  2. Japanese Society of Anesthesiologists. JSA airway management guideline 2014: to improve the safety of induction of anesthesia. J Anesth. 2014;28:482–93.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

HY and AK wrote the original draft. IF and ST helped in the conceptualization of this manuscript. YN contributed to the conceptualization, supervision, and review. All authors reviewed and approved the final manuscript.

Corresponding author

Correspondence to Haruyuki Yuasa.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

We obtained written informed consent from the patient to present this case.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kitaura, A., Fukuda, I., Yuasa, H. et al. Difficult ventilation due to an undetected mask crack. JA Clin Rep 9, 80 (2023). https://doi.org/10.1186/s40981-023-00672-2

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40981-023-00672-2