Skip to main content
  • Letter to the Editor
  • Open access
  • Published:

Failed ventilation via i-gel due to air leakage of the drainage port

To the Editor

Supraglottic airway devices (SADs) are commonly used to provide adequate ventilation in anesthesia practice. However, failed ventilation due to incomplete airway management via such devices has been reported, although the incidence is rare [1]. i-gel™ (i-gel, Intersurgical, Berkshire, UK) is a SAD equipped with a gel-like cuff made of thermoplastic elastomer that seals the laryngo-pharyngeal space [2], and has an additional small lumen for esophageal drainage. We herein report a case of failed ventilation because of massive air leakage from an i-gel drainage port. Written informed consent was obtained from the patient for publication of this case report and its accompanying images. A 74-year-old male (height, 163 cm; weight, 69 kg) underwent transurethral resection of bladder tumor under general anesthesia, induction of which was done with remifentanil 0.12 μg/kg/min, fentanyl 100 μg, and propofol 80 mg. Any unique anatomical features had never been pointed out to him. After obtaining muscular relaxation with rocuronium 60 mg, an i-gel #4 was inserted without any specific resistance; however, adequate ventilation was not achieved due to massive air leakage. We attempted to provide manual ventilation with approximately airway pressure of 15 cmH2O; however, the target airway pressure was not achieved because of air leakage and each expiratory volume was less than 50 ml. Several attempts were made to properly position the i-gel, but all failed. Close observation revealed that the location of the leakage was a drainage port of the i-gel. Covering the drainage port with a finger significantly reduced the air leakage, and adequate tidal volumes (450–500 ml with airway pressure of 15 cmH2O) were achieved. Subsequently, we covered the port with plastic tape, after which adequate airway seal was maintained during surgery.

When ventilating with an i-gel, its distal tip with the distal port of drainage duct is positioned at the esophageal inlet (Figs. 1A and 2A, B). However, in the present case, the distal tip was located at the laryngeal inlet (Figs. 1B and 3A, B). Thus, the inspired air through the ventilation port of the i-gel exited directly through the drainage port located at the laryngeal inlet and did not reach the lungs. An i-gel has a smaller drainage port as well as a smaller distal ventilation port compared to other SADs, which might have allowed the i-gel to migrate into the laryngeal inlet. i-gel might have caused failed ventilation in the current case due to this unique issue. When massive air leakage from the i-gel drainage port is observed, covering the drainage port may be a promising option.

Fig. 1
figure 1

A, B Illustrations of i-gel placement. A Adequate i-gel placement. Ventilated air flows in the direction of the trachea. B Inadequate i-gel placement, in which the drainage port at the distal tip is located at the laryngeal inlet. Ventilated air leaks out through the drainage port

Fig. 2
figure 2

A, B Bronchoscopic view of deal i-gel placement in a manikin. A A view from the distal tip of i-gel ventilation duct. Arytenoid cartilage and vocal cords can be observed. B A view from the distal tip of i-gel drainage duct, showing only the esophagus

Fig. 3
figure 3

A, B Bronchoscopic view of approximate i-gel placement in a manikin simulating the present case. A A view from the distal tip of i-gel ventilation duct. Vocal cords, but not the arytenoid cartilage, can be observed. B A view from the distal tip of i-gel drainage duct. Vocal cords and the edge of the arytenoid cartilage, but not the esophagus, can be observed

Availability of data and materials

Not applicable.

Abbreviations

SAD:

Supraglottic airway device

References

  1. Saito T, Liu W, Chew ST, Ti LK. Incidence of and risk factors for difficult ventilation via a supraglottic airway device in a population of 14,480 patients from South-East Asia. Anaesthesia. 2015;70:1079–83.

    Article  CAS  PubMed  Google Scholar 

  2. Levitan RM, Kinkle WC. Initial anatomic investigations of the i-gel airway: a novel supraglottic airway without inflatable cuff. Anaesthesia. 2005;60:1022–6.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors acknowledge the Scientific English Editing Section of Fukushima Medical University for their proofreading.

Funding

This study was supported solely by departmental funding from our institution.

Author information

Authors and Affiliations

Authors

Contributions

KK, MK, and SI designed the report. KK and SI wrote the first draft of the manuscript. SI revised the manuscript. All authors have read and approved the final version of the manuscript and agreed to its submission for publication.

Corresponding author

Correspondence to Satoki Inoue.

Ethics declarations

Ethics approval and consent to participate

The institutional review board waived the approval requirement of this report. Written informed consent for publication was obtained from the patient.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kuwana, K., Kobayashi, M. & Inoue, S. Failed ventilation via i-gel due to air leakage of the drainage port. JA Clin Rep 9, 62 (2023). https://doi.org/10.1186/s40981-023-00655-3

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40981-023-00655-3